Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke RPM 80 Inductive Pickup: (1 year) CAL VER DATE: 27-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 79 STANDARD: Philips PM 5133 STANDARD: Philips PM 3050 ============================================================================= # # Source: # Fluke 88 Service Manual (PN 880906 Dec. 1990, Rev. 2, 9/94). # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # RPM80.bmp # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 STD Philips PM 5133 1.003 STD Philips PM 3050 1.004 HEAD {INDUCTIVE PICKUP TEST} 1.005 DISP Solder a 10-inch piece of 14 AWG wire to each 1.005 DISP end of a 10 Ohm resistor. 1.006 PIC RPM80 1.007 DISP Set up the PM 5133 Function Generator as follows: 1.007 DISP [32] Mod/Sweep Period ..................... 10ms 1.007 DISP [32] Mode ...... Single (or N Cycle) 1.007 DISP [32] Range ....................... 2M 1.007 DISP [32] Waveform ................. Triangle 1.007 DISP [32] Start Phase .................... -pi/2 1.007 DISP [32] Period ....................... x1 1.007 DISP [32] Std By/Cont ..................... Cont 1.008 DISP Set up the PM 5133 Function Generator as follows: 1.008 DISP [32] 600 Ohm/50 Ohm ................... 50 Ohm 1.008 DISP [32] AM/FM Depth ...................... Off 1.008 DISP [32] Attenuation ..................... None 1.008 DISP [32] Duty Cycle ...................... Out 1.008 DISP [32] DC Offset ....................... In 1.008 DISP [32] Frequency ................. 0.666MHz 1.009 DISP Set up the PM 5030 Oscilloscope as follows: 1.009 DISP [32] Vertical Atten. ................. 0.5V/div 1.009 DISP [32] Horz. Time Base ................ 0.5us/div 1.009 DISP [32] Trigger Source ................ Channel 2 1.010 DISP Adjust the amplitude of the PM 5133 to produce a 3Vpp 1.010 DISP triangle wave. 1.011 DISP Set up the PM 5030 Oscilloscope as follows: 1.011 DISP [32] Vertical Atten. ................. 1.0V/div 1.011 DISP [32] Horz. Time Base ................ 5.0ms/div 1.011 DISP [32] Trigger Source ................ Channel 1 1.012 DISP Check that the peak voltage is greater than 6.0V 1.012 DISP and decays to less than 1.0V between pulses. 1.013 DISP Adjust the peak voltage to 6V using the variable voltage 1.013 DISP gain control (VAR). 1.014 DISP Set up the PM 5030 Oscilloscope as follows: 1.014 DISP [32] Horz. Time Base ................ 1.0ms/div 1.015 MATH MEM1 = 2.4 1.016 MEMI Enter the amplitude after 3.0ms: 1.017 MEME 1.018 MEMC V +0.5U -0.8U 2.001 DISP Set up the PM 5030 Oscilloscope as follows: 2.001 DISP [32] Horz Time Base ................ 5.0ms/div 2.001 DISP [32] Voltage Var Ctrl ...................... CAL 2.002 DISP Turn the inductive pickup so that "SPARK PLUG SIDE" 2.002 DISP points along the wire connected to the LOW output of 2.002 DISP the function generator. 2.003 EVAL Is the {waveform less than 2V}? 3.001 END